Quantum Light Instruments: Quantas-1064 Nd:YAG diode-pumped Q-switched laser

Quantas-1064 Nd:YAG diode-pumped Q-switched laser

  • Up to 32 mJ pulse energy at 1064 nm.
  • Air cooled – no liquids.
  • Short pulse duration < 8 ns.
  • Weight is < 5 kg incl. power supply.
  • Variable up to 50 Hz repetition rate.
  • Built-in sync pulse generator for triggering user equipment.
  • Remote control via built-in Ethernet or Wifi interface.
  • Optional 2nd, 3rd, 4th or 5th   harmonic generators.
  • Optional attenuator for fundamental, 2ndand 3rd  harmonics.
  • Optional fiber coupled output.
  • Guaranteed > 1 Gshot life time.

Quantas-1064 Nd:YAG diode-pumped Q-switched laser

Download product data sheet.

Description

Quantas series lasers are compact, fully air-cooled (no water), Q-switched lasers designed for wide range of applications that require low pulse repetition rate and high peak power pulses (Q1D model produces ~5 MW peak power). Due good thermal properties of Nd:YAG crystal Q1-1064 can operate at higher pulse repetition rates in comparison to Q1-1053. Laser delivers up to 30 mJ at 20 Hz repetition rate or up to 10 mJ at 50 Hz repetition rate. Typical applications are Light Induced Breakdown Spectroscopy (LIBS), Light Induced Fluorescence Spectroscopy (LIF), laser ablation, remote sensing and many others. Less than 8 ns pulse duration allows efficient fundamental wavelength conversion to higher harmonics with shortest wavelength available of 213 nm. Wavelength extensions into infrared range, by use of OPO, are available by request. Quantas is completely air cooled laser due its excellent wall-plug efficiency. Liquids are not used for heat transfer, as result, maintenance associated with regular replacements of cooling liquid and/or cleaning of cooling system is not required. Low jitter triggering pulses for user equipment are available with up to 300 µs (YAG models) or 450 µs (YLF models) lead in internal triggering mode. If required, laser pulsing can be externally triggered from delay generator, allowing operation in single-shot or variable pulse repetition modes. Laser controller has standard Ethernet (Wifi is optional) interface for convenient monitoring and control from personal computer.

Applications

  • Laser Induced Breakdown Spectroscopy (LIBS)
  • LCD repair
  • Time-of-Flight Spectroscopy (TOFS)
  • Laser Induced Fluorescence (LIF) Spectroscopy
  • Ophthalmology
  • Matrix Assisted Laser Desorption/Ionization (MALDI)
  • Remote sensing
  • Laser ablation

Specifications 

Model Quantas-1064
Q1B Q1C Q1D
Wavelength 1064 nm 2)
Pulse energy 3)
Fundamental 10 mJ 18/16 mJ 32/30 mJ
532 nm 5 mJ 9/8 mJ 16/15 mJ
355 nm 3 mJ 5.5/5 mJ 9.5/9 mJ
266 nm 1.5 mJ 3/2.5 mJ 5/4 mJ
213 nm 0.5 mJ 1.2/1 mJ 2/1.6 mJ

Pulse repetition rate

Min Single-shot
Max 10, 20 or 50 Hz 10 or 20 Hz 10 or 20 Hz
Pulse duration < 8 ns  5)
Pulse-to-pulse energy stability < 0.5 % rms 6)
Power drift ± 3% 7)
Beam profile Nearly TEM00 Bell-shaped, >75 % fit to Gaussian
Beam divergence 8) < 2 mrad < 1.5 mrad
Beam diameter 9)  1.5 mm 2 mm 2.5 mm
Polarization Linear, horizontal at 1064 nm
Optical jitter < 1 ns rms 10)
Dimensions
Laser head (W x L x H) 113 × 230 × 112 mm3
Harmonics generator (W x L x H)     113 × 242 × 112 mm3
Controller unit (W x L x H) 85 × 165 × 50 mm3
Power adapter (W x L x H) 50 × 125 × 32 mm3
Operation environment 15-30°C, 10-80% humidity (non-condensing), 90-230 VAC, 30 W, 47-63 Hz 1
Min Single-shot
Max 10, 20 or 50 Hz 10 or 20 Hz 10 or 20 Hz
Pulse duration < 8 ns  5)
Pulse-to-pulse energy stability < 0.5 % rms 6)
Power drift ± 3% 7)
Beam profile Nearly TEM00 Bell-shaped, >75 % fit to Gaussian
Beam divergence 8) < 2 mrad < 1.5 mrad
Beam diameter 9)  1.5 mm 2 mm 2.5 mm
Polarization Linear, horizontal at 1064 nm
Optical jitter < 1 ns rms 10)
Dimensions
Laser head (W x L x H) 113 × 230 × 112 mm3
Harmonics generator (W x L x H)     113 × 242 × 112 mm3
Controller unit (W x L x H) 85 × 165 × 50 mm3
Power adapter (W x L x H) 50 × 125 × 32 mm3
Operation environment 15-30°C, 10-80% humidity (non-condensing), 90-230 VAC, 30 W, 47-63 Hz 1

1) The parameters marked typical are not specifications. They are indications of typical performance and might vary unit-to-unit. Unless stated otherwise all specifications are measured at 1064 nm at max pulse repetition rate. 
2) 1053 nm output wavelength models are available. For detailed specifications please check here. 
3) First number is for 10 Hz, second – for 20 Hz pulse repetition rate. 
4) Factory-set pulse repetition rate is fixed at max repetition rate. Single-shot or variable pulse repetition rate is possible when laser is externally triggered. Please inquire for higher rep. rates. 
5) FWHM at 1064 nm. Shorter pulse durations are available by request. 
6) Averaged from 500 pulses. 
7) Over 8 hour period after 20 minutes of warm-up when ambient temperature variation is less than ±2 °C. 
8) Full angle measured at 1/e2 level. 
9) Beam diameter is measured 20 cm from laser output at 1/e2 level. 
10) In respect to Q-switch triggering pulse. 